FLB/TLB-2x-Scope V1.0 (KW 09/2006 - 03.03.2006) 7
HOLD:
The FLB-CON2-2X operates with minimum scan times in the range of 100µs. For this reason
most of the SPCs that are connected at the digital error output TOLOUT have difficulties with
the safe detection of the resulting short changes of switching states.
By activating the respective HOLD selection button a pulse lengthening at the digital output of
the FLB-CON2-2X of up to 100 ms can be set.
DATA:
Displays the current averaged value of the sensor frontend. This value is also visualised in the
graphic display window with the corresponding reference windows REF1 ± TOL1 and REF2 ±
TOL2.
SCALE setting:
These selection buttons are used for setting the scaling type of the y-axis.
FIX: Fixed scaling of the y-axis (0...255 – results from 8-bit A/D conversion)
AUTO: Automatic adaptation of y-axis scaling to the current measured values
(zoom function).
Inspect button:
This button fixes the current screen of the display window. A dialog box then opens and offers
different possibilities for further processing of the display window under Windows. The
individual function elements of the inspect button are excellently suited for documentation
purposes.
Print: Prints the current display window.
Copy Page: Copies the current display window to the Windows clipboard, from where
it can be inserted and further processed in various Windows programs
(documentation in the word processor, graphics software, spreadsheet, ...).
Zoom: For zooming in or zooming out the display window.